The Martel LC-110 / LC-110H Calibrators are designed to take the loop calibrator class to the next level. The new LC-110 series features
a user friendly interface with dedicated buttons and a rotary encoder (Quick-Set Knob). This combination dramatically reduces the time
it takes to measure, or source voltage or current and power up a loop. The rugged case is contoured to easily fit a technician's hand
and the large back lit graphics LCD is best in class.
The LC-110H differs from the standard LC-110 in that it incorporates HART communications and supports a select set of the HART universal
and common practice commands. This unique feature allows the LC-110H to be used as both a loop calibrator and communication tool.
In the communicator mode the user will be able to read basic device information, perform diagnostic tests, and trim the calibration on
most HART enabled transmitters. In the past, this could only be done with a dedicated communicator, high-end multifunction calibrator
costing thousands of dollars, or a laptop computer with HART modem. The LC-110H will allow many more technicians to service and support
HART devices.
In addition to HART communications, we also gave the user the ability to get information out of the LC-110H. Need to quickly document
the parameters of all the HART transmitters in your plant? Just add the BetaLOG HART software/cable to capture and upload to twenty
configurations in either (.csv) or (.txt) format.
The data log tool features selectable capture interval from 1 to 60 seconds and a logging capacity of 9800 records or 99 individual
sessions. Each data sample contains the LC110H measurement, all four process variables, and the standard status conditions.
Ships complete with:
- NIST traceable calibration certificate with data
- Batteries
- Test leads
- Manual
Kit option includes:
- LC-110H loop calibrator with HART communications
- BetaLOG-HART data logging software
- USB communications cable
- Soft carrying case
- Mini-grabber® test leads
- Alligator clip test leads