Flocat LA10-A Volumetric Gas Flow Meter
Differential pressure flow meter for clean gases. Flow: 0.025 ml/min to 1500 l/min.
Overview
Features
- Microprocessor controlled
- Accuracy ±1% FS within stated ranges
- Variety of media choices
- Large selection of flow ranges
- Choice of output
- Turndown ratio 100:1
- Switch selectable gas calibrations
- Accuracy ±0.5% option
- Remote digital display available
- Totalizing mode available
- NIST Traceable Calibration Available
- CE Certified
Description
The Flocat LA10-A gas flow meter uses two of the best-studied physical properties of gases to measure flow: Pressure and Viscosity. Differential pressure measurement across a laminar flow element results in a gas flow meter that is inherently linear. Flocat LA10-A gas flow meters measure within the laminar region to achieve turndowns of 100:1 typically. A very wide range of measurement, 0.5 milliliter per minute full-scale to 1500 liters per minute full-scale, is standard. Unique laminar flow elements, along with modern electronics, ensure that each Flocat LA10-A gas flow meter is both affordable and accurate.
The Flocat LA10-A gas flow meter is a simple yet rugged device. The membrane switch user interface and dynamically labeled graphic display combine to provide exceedingly flexible and user-friendly operation. The Flocat LA10-A gas flow meter can have several screen "modes" depending on how the device is ordered.
Volumetric Flocat LA10-A gas flow meters are intended for use in low pressure applications. This is because an accurate measurement of the volumetric flow rate by means of differential pressure requires the flow at the differential pressure sensor to be in a laminar state.
The Flocat LA10-A is also available with flow control capability.
Applications for the Flocat LA10-A Gas Flow Meter
- Semiconductor Manufacture
- Pollution Monitoring
- Gas Blending
- Gas Chromatography
- Chemical Reactions
Operating Limits of the Flocat LA10-A Gas Flow Meter
- Pressure: 145 PSIG Max
- Temperature: -10 to 50°C (operating)
Specifications for the Flocat LA10-A Gas Flow Meter
- Accuracy
- ±1% FS
- Repeatability
- ±0.5% FS
- Turndown Ratio
- 100:1
- Response Time
- 10 ms
- Operating Temperature
- -10 to +50°C
- Zero Shift*
- 0.02% FS/°C/atm
- Span Shift*
- 0.02% FS/°C/atm
- Measurable Flow Rate
- 128% FS
- Pressure
- 145 PSIG Max
- Supply Voltage
- 7-30 VDC
- Supply Current
- 30mA
- Output
- 0-5 VDC, 0-10 VDC, 4-20 mA
- Wetted Surfaces
- 303 & 302 SS, Viton, silicone RTV, glass reinforced nylon
- Connections
- NPT (F)
- Media
- Air, Argon, Nitrogen, Oxygen, Hydrogen, Helium, Carbon Dioxide, Methane, Propane, Carbon Monoxide, Ethane, Nitrous Oxide, Neon**, Nitric Oxide
*For full terminology discussion, consult Instrumart
**Consult Instrumart for details on switch set-up
Need Help? Call a Flow engineer at 1-800-884-4967
We're open Mo-Th 8am to 5:30pm. Fr 8am to 5pm ET