Fluke - Hart Scientific 1551A and 1552A "Stik" Thermometers
Highly precise and durable digital replacement for mercury-in-glass thermometers
Availability
Usually ships in 2-3 weeks
Features
- Manufactured by Fluke / Hart Scientific
- Accuracy of ±0.05°C (±0.09°F) over full range
- Intrinsically safe (ATEX and IECEx compliant)
- Two ranges from which to choose: -50 to 160°C or -80 to 300°C
- User-configurable temperature trend/stability indicator
- Display temperature in °C or °F
- Optional datalogging to internal memory
- 300-hour battery life
- Percent battery-life and low-battery indicator
- NVLAP-accredited, NIST-traceable calibration included
Includes FREE
The Fluke-Hart Scientific 1551A and 1552A "Stik" thermometers provide a highly precise and durable digital replacement for mercury-in-glass thermometers. They are intrinsically safe (ATEX and IECEx compliant), so they can be used safely where potentially explosive gases may be present. More durable than mercury-in-glass thermometers, they can be used reliably outdoors or on the production floor.
The stainless steel probe and digital readout of the Fluke-Hart Scientific 1551A and 1152A "Stik" thermometers are fixed together and calibrated as a system. Both models are intrinsically safe and accurate to ±0.05°C over their full range. The models are differentiated only by the sensing element and temperature range. The 1551A features a robust thin-film RTD sensor with ranges from –50 to 160°C and primarily targets the petroleum industry. The 1552A utilizes a wire-wound PRT sensor with ranges from –80 to 300 °C and targets the pharmaceutical industry. Of course many other industries have applications that require an easy-to-use, stable and portable reference thermometer and the "Stik" thermometer is the perfect solution.
The Fluke-Hart Scientific 1551A and 1552A "Stik" thermometers include a user-configurable stability/trend indicator that lets users know when temperature is stable enough to record an accurate measurement. A user-settable auto-off function extends typical battery life to three hundred hours while a low-battery indicator and stop-measure function prevent erroneous measurements from being made due to low battery life. A simple three-point calibration function allows users to easily and accurately calibrate the 1551A and 1552A. Datalogging up to 10,000 readings to an internal memory of is available as an option. The large backlit LCD display rotates 90 degrees, making it easy to read from any angle.
Fluke – Hart Scientific 1551A and 1552A "Stik" Thermometers Specifications:
| Parameter | 1551A | 1552A |
|---|---|---|
| Temperature Range | -50 to 160°C | -80 to 300°C |
| Accuracy | ±0.05°C | ±0.05°C |
| Sensor Type | 100 Ω thin-film RTD | 100 Ω wire-wound PRT |
| Temperature Coefficient | 0.00385 Ω/Ω/°C nominal | 0.00385 Ω/Ω/°C nominal |
| Sensor Length | ≤ 10 mm | ≤ 30 mm |
| Sheath Material | Stainless steel | Stainless steel |
| Response Time | 4.8 mm diameter probe (3/16 in): 14 seconds 6.35 mm diameter probe (1/4 in): 21 seconds |
|
| Probe Hysteresis | ±0.01°C | ±0.01°C |
| Resolution | Selectable: 0.1, 0.01 (default), 0.001 | |
| Sample Rate | Selectable: 0.5 seconds, 1 second (default), 2 seconds | |
| Datalogging (optional) | Up to 10,000 time-stamped readings stored to internal memory | |
| Logging Intervals | 2 seconds to 60 minutes | |
| Damping | Moving average of the most recent 2 to 10 readings | |
| Communications | RS-232 stereo jack (access calibration parameters only) | |
| EMC Compliance | EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class "B" |
|
| Enclosure Rating | IP50 | |
| Calibration (Included) | NVLAP-accredited, NIST-traceable | |