GE Druck UPS-II Loop Calibrator
$532.00

Availability

Call for availability

Need Help? Call us M-F 8 AM to 8 PM ET 1-800-884-4967

Features

  • Part number UPS-II
  • Manufactured by GE Druck
  • Sources and reads milliamps
  • Simulates 2-wire transmitters
  • Auto-ramp cycling for endurance tests
  • Auto-stepping for hands-free calibration
  • Powers and reads 2-wire transmitters
  • Reads and simulates dp-transmitters in percentage of flow
  • Fixed calibration currents
  • Fixed currents for valve stroking
  • Output Range: 0 - 21.00mA
  • Input Range: 0 - 21.00mA
  • Accuracy: 0.05% of Range
  • Resolution: 10mA
  • Internal Loop Power: 24 VDC Stabilized
  • External Loop Power: 56 VDC Max
  • Intrinsically Safe version available

The GE Druck UPS-II loop calibrator is an advanced handheld process calibrator tool. The GE Druck UPS-II loop calibrator can power (24 VDC) and read 2-wire transmitters to perform field calibrations. Output currents are adjustable to a resolution of 10 microamps in both source and transmitter simulation modes.

Fixed Steps

The GE Druck UPS-II loop calibrator can source or stimulate fixed calibration currents to test or calibrate receivers and to stroke valves. Different current levels are available from the menu.

4-20mA Linear 4 - 8 - 12 - 16 - 20mA
0-20mA Linear 0 - 5 - 10 - 15 - 20mA
4-20mA sq.rt (Flow) 4 - 5 - 8 - 13 - 20mA
0-20mA sq.rt 0 - 1.25 - 5 - 11.25 - 20mA
4-20mA 3.8 - 4 - 4.2 - 12 - 19 - 20 - 21mA

The GE Druck UPS-II loop calibrator's output currents can be changed manually or automatically. The time between steps is 10 seconds in the auto-stepping mode. In the valve mode, auto-stepping will be 3.8 - 4 – 4.2 or 19 - 20 - 21mA.

Ramping

The GE Druck UPS-II loop calibrator can source or simulate with high resolution a continuous up / down cycle between 4 and 20mA or 0 and 20mA. The travel time between limits is 60 seconds.

Reading in Percentage

The GE Druck UPS-II loop calibrator can display readings in percentage of a 4-20mA or 0-20mA span and can also read or simulate dP-transmitters directly in percentage of flow.

The standard instrument kit provided with the GE Druck UPS-II loop calibrator includes carrying case, alkaline batteries, set of test leads, users guide, and calibration certificate.

An intrinsically safe version of the GE Druck UPS-II loop calibrator is available.

Power Supplies
GE Druck Charger/ Power Supply
For UPS-II Current Loop Calibrators
Starting at
$59.89
Cases
Carrying cases, spare cases
Pelican Cases
Watertight, crushproof and dustproof cases for the ultimate in protection
Starting at
$44.00